Home    中文  
 
  • Search
  • lucene Search
  • Citation
  • Fig/Tab
  • Adv Search
Just Accepted  |  Current Issue  |  Archive  |  Featured Articles  |  Most Read  |  Most Download  |  Most Cited

Chinese Journal of Interventional Radiology(Electronic Edition) ›› 2018, Vol. 06 ›› Issue (02): 154-157. doi: 10.3877/cma.j.issn.2095-5782.2018.02.013

Special Issue:

• Fundamental Research • Previous Articles     Next Articles

Calculation of radial dose distribution of 125I radioactive particle chain by 3D treatment planning system

Dechao Jiao1, Zhihui Zhong2, Rongfang Niu1, Xinwei Han1,(), Bo Ma1, Jianzhuang Ren1, Zhan'guo Sun1, Juanfang Liu1, Jin Shi1   

  1. 1. The First Affiliated Hospital of Zhengzhou University, Zhengzhou 450052, China
    2. The Tumor Hospital Affiliated Sun yet-sen University, Guangzhou 510050, China
  • Received:2017-09-26 Online:2018-05-01 Published:2018-05-01
  • Contact: Xinwei Han
  • About author:
    Corresponding author: Han Xinwei, Email:

Abstract:

Objective:

To calculate the radial dose distribution of 125I radioactive particle chain with different lengths, distances and activities using the three dimensional treatment planning system (3D-TPS) .

Methods:

The single, double and three 125I radioactive particle chains were implanted into the equivalent human body membrane and scanned by MSCT. The radial dose distribution of particle chains was calculated on 3D-TPS after the scanned images were imported.

Results:

The dose field of 125I radioactive particle chain was cylindrical, which dropped fast if the distance was more than 1 cm from the 125I chain. The regression equations of single, double and three 125I cumulative dose (dose, Gy) , seed activity (activity, mCi) and distance from the center (cm) were: ln dose=43.3 activity-24.2 distance+48.4; ln dose=95.4 activity-55.8 distance+112.0; ln dose=138.0 activity-79.8 distance+160.5.

Conclusion:

The dose field of 125I radioactive particle chain is mainly affected by the seed activity and radial distance from center, while the affection of 125I radioactive particle chain length is weak. The measurement results can provide reference for clinical application.

Key words: 3D treatment planning system, Computer aided system, Particle chain, Radiation dose